Description: |
Analysis system SIEMENS
used machine
Siemens Advia Centaur XP Analysis System
With Siemens Healthineers services and maintenance until the last
minute 02/2022
Modern AE immunoassay technology
Meet clinical requirements with acridinium ester (AE) technology,
which ensures the necessary flexibility and sensitivity for new and
innovative assays.
Comprehensive diagnostic clarification of various clinical pictures
thanks to a test menu with over 70 different assays.
Simplified inventory management and lower costs by using the same
ready-to-use reagents on all ADVIA Centaur analysis systems.
Simplified operation, continuous operation
All samples, reagents and consumables can be loaded and unloaded
during operation.
Easier training and operation thanks to an intuitive, icon-based user
interface.
Fewer operator intervention because daily maintenance is low and the
system does not need to be cleaned monthly or booted up daily.
New and updated test definitions can be read by the customer by simply
scanning the 2D barcodes contained in the instructions for use.
Consistent, predictable turnaround time (TAT) and accuracy
Up to 240 tests/hour for a short turnaround time even at peak
times.
Rapid prioritization of emergency samples possible at any time thanks
to a dedicated access (STAT port).
Disposable pipetting tips prevent carry-over and contamination of the
samples.
Automatically perform repeat and confirmatory tests for reactive
hepatitis B surface antigen samples using Siemens Healthineers SMART
algorithms.
Direct connection option
Adapt to different sample volumes through easy connection to Aptio and
VersaCell® X3 automation solutions.
Efficient operation thanks to diagnostic IT including CentraLink®
data management system
Atellica Process Manager and Smart Remote Services for Diagnostics
Direct pipetting from the automation solution through point-in-space
technology.
Shorter handling times for emergency samples thanks to a dedicated
STAT port that allows emergency samples to be loaded from the front
even when connected to an automation solution.
Technical Specifications
Specifications
System Type: Random Access Immunoassay System
Throughput: Up to 240 tests per hour
Time until the first test result
18 minutes, further results in a 15-second rhythm (depending on the
method)
Method capacity on board: 30 test methods
Indication groups§§
Allergy, anemia, diabetes, fertility, hepatitis, HIV,
immunosuppressants, infection serology, cardiology,
Bone metabolism, liver fibrosis, adrenal function, oncology, thyroid
diseases, sepsis, therapeutic drug monitoring (TDM)
Continuous operation
Loading and unloading of samples and consumables at any time during
operation. This also includes sample racks, emergency samples,
reagents, additional reagents, reaction cuvettes, sample tips,
calibrators, controls, washing solutions, water and waste.
Samples and Sample Management: Sample Tubes
3 ml, 5 ml, 7 ml and 10 ml tubes, 1 ml and 2 ml secondary vessels,
microvessels
Sample feed
Universal rack with 5 positions for different tube types, total
capacity 180 samples; Loading and unloading during operation
Validated sample types
Serum, plasma, urine, whole blood (depending on the test method)
Sample volume per test
10–200 μl, depending on the test method
Sample Integrity Testing
Pressure sensor technology with clot detection and management, sample
volume monitoring including small quantity detection and management;
each with marking and documentation
Emergency Sample Management
Emergency samples can be supplied at any time via a dedicated access
(STAT port); Emergency samples are analysed as a matter of priority
Automatic test repetition: Customizable
Sample dilution
Automatic further dilution (up to 1:2500), depending on the test
method
Automatic reflex testing
Automatic execution of additional tests, depending on the results of
the first test
Pipetting tips
Capacity sufficient for 840 disposable sample tips
Protection against carry-over between samples
Disposable pipetting tips prevent carry-over between samples
Barcodes
Sample Barcodes
Up to 20 characters, Interleaved 2 of 5, Code 39, Code 128, Codabar,
A, B and special characters (.-+/$%)
Response Range
Reaction cuvettes
Capacity sufficient for 1,000 optically high-quality plastic
cuvettes
Reaction environment: Air-tempered, 37 °C
Calculation of the test result
Test repetition and test response confirmation using SMART algorithm
software, result output after completion
Assay Technology
Chemiluminescence with modern acridinium ester technology
Reagent management
Reagent management
Reagent compartment with 30 positions, cooled (between 4°C and
10°C), automatic mixing of ReadyPacks
Onboard reagent capacity: 30 tests
Reagent containers
ADVIA Centaur® ReadyPack® (different pack K41227004 sizes
available)
Reagent Integrity Check
Reagent identification by barcode, automatic inventory recording,
recording of the validity of calibrations, recording of reagent
stability in the system, each including marking and documentation,
reporting of expired/low reagent stocks
Onboard stability
28 days (on average)
Mixing the reagents
ADVIA Centaur ReadyPacks are automatically kept in motion in the
system
Additional reagents
Additional reagent compartment with 25 positions, cooled (4–8°C)
Calibration/Quality Control (QC)
Auto Calibration/Auto QC
Automatic QC, optionally according to number of tests, control, date
and time
QC data
Powerful QC package with Levey Jennings curves, Westgard rules and
RiliBÄK support
User Interface/Data Management
Monitor
22-inch LCD touch screen monitor with height adjustment
Operating System: Microsoft Windows 7
System Documentation
Operator's manual, quick start guide and context-sensitive online
help
Memory
1,000,000 test results/QC results, backup to removable media
possible
Onboard Maintenance Logs And Host Interface
RS232C bidirectional with host query, Ethernet port available
Host Query: And
Remote access and maintenance
1000BASE-T Ethernet port for remote access and modem for remote
maintenance from Siemens Healthineers with Smart Remote Services
Removable media: CD, DVD or USB stick
General technical data
Energy supply
200–240 V at 50/60 Hz; 1.2 kW
Maximum water consumption: 2.5 liters per hour
Wastewater Requirements
At least 2.5 liters per hour
Dimensions
With monitor: 167 x 196 x 104 cm (H x W x D)
Without monitor: 133 x 196 x 104 cm (H x W x D)
Weight: 585 kg
000
Conformance
Complies with international environmental, health and safety
standards
Noise level:max. 65dB
Heat dissipation during operation: 5,100 BTU/hour
Ambient temperature:18–30°C
Humidity
20–80%, non-condensing; both ambient temperature and humidity are
recorded by the system
Load capacity of the floor: 412kg/ m² seller offer No. resale
24568
|